Accuracy
- 0.25 % of Sensor Range (Including Linearity, Hysteresis & Repeatability)
Approvals & Certifications
- ATEX Intrinsically Safe II 1 G EEx ia IIC T4
- CE
Approvals & Certifications
- Calibration Certificate
- Inspection Certificate
Cable Material
- Cable Gland: Butyl Rubber
- Cable Jacket: High-Density Polyethylene (HDPE)
Compensated Temperature
- 0° to 80° C (32° to 176° F)
Enclosure, Body Material
- Corrosion Resistant 316 Stainless Steel
Materials of Construction
- Self-Flushing Nose Cone: PVC for 510LT
Operating Temperature
- -20° to 80° C (-4° to 176° F)
Overpressure & Static Pressure Limits
- Overpressure Protection: 1.5 times Span
Pressure Range
- bar: 0 to 13.8
- Feet WC: 0 to 463
- kPa: 0 to 1,380
- psi: 0 to 200
Stability
- Long Term: ± 0.1 % of Sensor Range Per Year
Storage Temperature
- -40° to 80° C (-40° to 176° F)
Supply Voltage
- 10-32 Vdc
- Instrinsically Safe: 28 Vdc, Max.
Temperature Effect
- Span ≤ 1 % of Range Per 38° C (100° F), Calibrated at 25° C (77° F)
Zero Offset
- Zero Offset/Span Offset: ± 0.5 % of Range at 25° C (77° F)