A15/82 Fluoride Monitor

Brand: Analytical Technology Inc

ATI’s Model A15/82 Fluoride Monitor provides continuous measurement of free fluoride concentration in potable water. 

The system employs a fluoride sensitive ion selective electrode (ISE), which provides reliable measurements down to 0.1 PPM and as high as 1000 PPM. 

A chemistry module provides sample conditioning for the sensor, and the measured fluoride concentration is displayed on a separate electronics module that also provides alarm and Analog output functions.

Fluoride ISE sensors measure F-ion in solution the same way that a pH sensor measures hydrogen ions. 

A lanthanum fluoride crystal on the tip of the sensor develops a voltage that is proportional to fluoride ion activity.

An integral reference electrode is used as a comparator to measure the developed voltage, with the measurement made at a differential input amplifier. 

Since the activity of fluoride ions in solution is a function of pH and ionic strength, a small amount of buffer solution is added to the measured sample. 

This creates a stable condition in which the concentration of fluoride ion and the activity of fluoride ion are directly proportional.

The A15/82 Fluoride Monitor requires very little maintenance. 

One gallon of buffer will operate the system for approximately 45 days. 

The system also performs an automatic calibration routine to correct for electrode “zero” drift. 

The fluoride standard is automatically pumped into the chemistry module and the analyzer zero is automatically adjusted. 

The frequency of this calibration is user-selectable from every 1 to 999 hours. 

Calibration standards are easily mixed from a stock solution provided with the unit, and the buffer is prepared using common vinegar.

Specifications

Range

  • 0-1.00 ppm Min., 0-1000 ppm Max.

Accuracy

  • ± 5 % of span

Repeatability

  • ± 2 % of span

Drift

  • < 0.1 ppm per month

Display

  • 16 Character Alphanumeric Backlit LCD

Control Relays

  • Two SPDT Relays, 5A @ 220 Vac Resistive.
  • Programmable Deadband and Time Delay.

Control Mode

  • On/Off with Variable Deadband and Time Delay

Alarm Output

  • Independent Alarm Relay (SPDT) with Dual Setpoint.

Analog Output

  • Isolated 4-20 mA, 600 ohm Max. Load. Programmable Output Span. Output may be Inverted.

Operating Conditions

  • 0-50° C., 0-95 % R.H. Non-Condensing.

Power

  • 110/220 Vac ± 10 %, 50/60 Hz

Enclosure

  • Panel Mount Standard, NEMA 4X Wall Mount Optional