Q22PX pH / ORP Universal Sensor

Part Number: Q22-P1-2-1-1-2 Brand: Analytical Technology Inc

Universal PH Sensor

Price
$740.00USD
Stock
7 In Stock

ATi’s Q22 specialty pH/ORP universal sensors are designed for use with all non-ATi analyzer systems and generic panel meters, allowing end-users to utilize the same robust features of the standard ATi Q25 sensors.

ATi’s Q22 PX universal or loop powered sensors emulate standard, non-amplified combination sensors and are designed to interface with all non-ATi pH/ORP analyzers.

The universal sensors require no power supply as the preamplifier is powered by an internal lithium battery.

The pH sensor outputs a mV signal in the range of either -500 to +500 mV or 0 to 2 VDC; both are linear over the 0 to 14 pH range.

The ORP (Oxidation Reduction Potential) sensor outputs an mV signal that ranges from -1000 to +1000 mV.

These sensors may be used as replacement sensors for use with all non-ATi analyzers.

The Loop-powered sensors require an external 7-30 VDC power supply and provide a 4-20 mA output signal for direct connection to generic panel meters, data acquisition systems, or other devices that accept a current input.

The sensor includes integral temperature compensation of the pH signal, so no temperature output is provided.

The sensor output is uncalibrated, therefore, the receiving instrument must be capable of providing user calibration of the pH or ORP signal on a periodic basis.
Features
  • Differential-Style Sensor
  • PEEK® or 316SS Sensor Body Construction
  • Replaceable Sensor Saltbridge 
  • Integral Preamplifier
  • Multiple Electrode Types

Specifications

Measuring Range

  • 0 to 14 pH
  • 0 to +1000 mV or ± 1000 mV

Sensitivity

  • 0.002 pH
  • 0.2 mV

Stability

  • 0.02 pH
  • 2 mV (per 24 Hours, Non-Cumulative)

Wetted Materials

  • PEEK
  • Ceramic
  • Titanium
  • Viton
  • EDPM
  • Platinum or Gold : ORP
  • 316 Stainless Steel with Sanitary or Insertion Body Styles

Temperature Compensation

  • Pt100 or Pt1000 RTD

Sensor Cable

  • 6 Conductor Plus 2 Shields, HDPE Jacket

Temperature Range

  • -5° to 95° C (23° to 203ºC)

Pressure Range

  • 0 to 100 psig

Maximum Flow Rate

  • 10 Feet (3 Meters) Per Second

Maximum Sensor to Analyzer Distance

  • 3000 Feet (914 Meters): Universal
  • 1000 Feet (305 Meters): Loop Powered

Sensor Body Options

  • 1” NPT Convertible
  • 1 1/4” Insertion
  • 1-1/2” or 2” Sanitary-Style

Shipping Weight

  • 1 Pound (0.45 kg)

show more